AvCount Lube Low-Visc Particle Counter - SA1910-0

AvCount Lube Low-Visc Particle Counter – SA1910-0

The AvCount Lube Low-Visc particle counter determines particle concentration in oils with a viscosity up to 64 mm2/s. FFKM seals have been fitted to resist aggressive oils.

Simple operation
Ideal for handling lubricating and hydraulic oils
3 minute test time
Programmable via PC

With a simple test procedure and user friendly features, operator skill or expertise is kept to a minimum.

AvCount Lube Low Visc can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, AS4059 cleanliness codes.

Supplied with: sample probe, waste pipe, universal voltage power supplies, USB cable, calibration certificate and USB memory stick containing handbooks and ProTrend software for test method programming and data download.

Methods

ASTM D6876, ASTM D7619, ASTM D7647, IP 565GJB 420-1987/420-A-1996/420B-2006, IEC 60970, ISO 4406, ISO 60970

Features

  • Calibrated to ISO 11171, traceable to NIST SRM 2806d
  • Supplied with ASTM D7647, ISO 60970 and ISO 4406 loaded as standard
  • FFKM seals compatible with synthetic oils
  • Suitable for fluids up to 64 mm2/s viscosity
  • Precise and reliable results
  • Ultra compact rugged, steel case suitable for portable use
  • Data storage for up to 600 measurements
  • Internal Double Pump System
  • USB connection for PC connectivity

Details

Calibration & Verification
AvCount Lite utilises 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorised regional laboratories. Seta authorised training is required to perform a calibration. Verification can be performed using AvCount Verification Material.

ProTrend (SA1810-0) is supplied with the instrument and can report results to ISO 4406, NAS 1638 and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.

Specifications

Particle Size Range4 – 100 µm(c) (in accordance with ISO 11171)
Number of Size BandsUp to 16 when connected to Protrend, 6 when used as a stand-alone
Maximum Particle Count Per ml60 000 particles/ml
Sample Viscosity(Max) 64 mm2/s
Sample Temperature Range0 – 70 °C
Results600 internal, unlimited on PC
Sample Flow Rate30 ml/min ±5 ml/min
Total Sample Volume Used80 ml (includes rinse cycles) for ASTM D7647, 20 ml other methods
Power Supply100 – 240 Vac, 50/60 Hz
Size (HxWxD)41027 mm × 32 mm × 28 mmx 560 x 640 mm
Weight7 kg