AvCount3 Particle Counter for Skydrol – SA1105-0
The AvCount3 for Skydrol is a compact bench-top automatic particle counter, specifically designed to measure the particulate of Skydrol hydraulic fluid.
Skydrol is an advanced fire resistant aviation hydraulic fluid which inhibits corrosion and prevents erosion damage to servo valves. It has been approved by most airframe manufacturers including Airbus, Boeing and BAE Systems and has been used in their products for over 40 years.
Due to the highly corrosive nature of Skydrol material this special version of the AvCount3 is fitted with Isolast seals and a protective inner case.
Methods
ASTM D975, ASTM D6786, ASTM D7619, ASTM D7647, IP 565, IP 630, DEF STAN 91-091, DEF STAN 91-86GOST 17216, ISO 4406, ISO 60970, JIS B 9932:2012, JIS B 9933:2021, JIS B 9934:2012NAS 1638
Features
- ISO 11171 calibration
- Cumulative counts/ml
- ISO 4406 cleanliness codes
- Colour touch screen
- Dilution ratio calculation
- Real time display of test progress
- User programmable
- LIMS, network and VNC connectivity
- Programmable alarm limits
- 14 embedded test methods
- In-field verification and calibration
- FFKM seals, compatible with synthetic oils
Details
Featuring a simple user interface with touch screen display, the fully-automated test means extensive operator training is not required before using the instrument. The simplicity of test set up and automation, gives operators the time to work on something else, helping to reduce labour costs.The AvCount3’s enhanced functionality includes a dilution protocol for testing of high viscosity samples, real-time display of test progess, 14 embedded test methods, user and sample identification and password protected levels.Statistical Quality Control software allows analysis of results in accordance with ASTM D6299.Supplied with: sample bottles, sample probe, waste pipe, universal voltage power supply, calibration certificate and instruction manual.
Specifications
Particle Size Range | ISO 11171: 4 µm(c) to 70 µm(c) (calibration for larger sizes available on request) |
Number of Measuring Channels | 16 size channels displayed on instrument, 4 µm(c) to 70 µm(c) and 2 µm(c) to 100 µm© (ISO 4402 sizes) |
Counts Per Measurement (Max) | 60,000 per ml |
Sample Viscosity (Max) | 64 mm2/s (using internal pump), 200 mm2/s (pressure fed @ 3 BarG) (SA1950-0 Sample Delivery System is available as an accessory) |
Sample Temperature Range | Ambient 0 to 70 °C |
Results | 500,000 result memory. Print via internal printer, export to LIMS, USB or QR code |
Sample Flow Rate | 30 ml/min ±5 ml/min |
Total Sample Volume Used | 80 ml for ASTM D7619 & IP 565, from 20 ml for other methods (includes flush cycles) |
Power | 100/240 V, 50/60 Hz, Auto-sensing universal power supply |
Size (HxWxD) | 37 cm x 23 cm x 27 cm |
Weight | 6 kg |