Clora R, with Autosampler, Standard Cup - 403747-01AS

Clora R, with Autosampler, Standard Cup – 403747-01AS

Clora R benchtops are compliant with ASTM D7536 and D4929 standards, and deliver a limit of detection of 0.13 ppm and a dynamic range up to 4wt%. Clora R is powered by MWDXRF, the same technology found in our signature sulfur analyzer, Sindie. This direct measurement technique does not require gasses or high temperature processes equating to easy operation and minimum maintenance requirements.

Clora R offers many features such as autosampler and Accu-flow.

Accu-flow technology eliminates particle settling and delivers results that better reflect crude streams as they exist in the refinery. Over a typical measurement cycle, the heavier particles can settle to the bottom of the sample cup and cause higher than normal results.

Methods

ASTM D7536, ASTM D4929

Features

Chlorine Analysis in Liquid Hydrocarbon and Aqueous Solutions

Clora R, with Autosampler, Standard Cup

  • Easy to use
  • Extremely low maintenance
  • Total chlorine analysis
  • For use in multiple application areas
  • Hassle-free sample preparation

Details

Easy to use
8-position carousel. LIMS integration for data management and transfer, intuitive 10-inch touch screen, measurement time 10-999 s, barcode reader auto fills sample name, USB connectivity, storage capacity for more than 50,000 results, supports up to 30 calibration curves.

Extremely low maintenance
No gasses, columns, heating elements, or quartz tubing.

Total chlorine analysis
Petroleum products, biofuels, aromatics and other chemicals, and water.

For use in multiple application areas
Refineries, petrochemical and additive plants, pipeline terminals, and test laboratories.

Hassle-free sample preparation
Utilizes standard XRF sample cups.

Specifications

Accu-flow:No
Auto Sampler:Yes
Calibration curves:Up to 30 calibration curves. Automatic and Manual Calibration functionality.
Catalysts:No
Communication Capabilities:USB, Ethernet; LIMS Connectivity
Consumable/Utility Gas:Internal Vacuum Pump
Dimensions: Depth:21 in / 54 cm
Dimensions: Height:16.5 in / 42 cm
Dimensions: Width:15.8 in / 40 cm
Element:Chlorine / 0.13 ppm
Measurement Time:User selectable 10 – 999 s, 300 – 600 s typical
Methods:ASTM D7536, ASTM D4929
Options:USB Printer (Receipt or Full Page), Consumables, Calibration Standards, Spare Parts.
Range:0.13 ppm – 4%
Repeatability:ppm – r

Xylene
0.05 – 0.15
1 – 0.17
5 – 0.31
10 – 0.50

Crude Oil
5 – 0.4
10 – 0.8
50 – 1.2

Water
5 – 0.6
10 – 1.0
50 – 1.5
Reproducibility:ppm – R

Xylene
0.05 – 0.27
1 – 0.29
5 – 0.53
10 – 0.90

Crude Oil
5 – 0.7
10 – 1.4
50 – 2.0

Water
5 – 1.0
10 – 1.5
50 – 2.5
Sample Cup Type:Standard XRF Cup
Sample Matrix:Hydrocarbon/ water
Specific Technology:MWDXRF
Type:Benchtop
Weight:75 lbs / 34 kg
What’s included?:Quick-Start Guide, Operations Manual, radiation survey, power cord, 6′ drain tubing, spare fan filter, spare fuse, 100 sample cups, 100 sample films, sample stand, spare primary window and 1 calibration set; 0-500 ppm Cl in Mineral Oil (6 x 10 mL). Printer optional.
XRF Film Type:Etnom

Spares and Accessories