PM-93 Pensky-Martens Flash Point Tester - 35000-0

PM-93 Pensky-Martens Flash Point Tester – 35000-0

The Seta PM-93 is a fully automated Pensky-Martens closed cup flash point tester which combines strict method conformance with the latest control technology and safety systems to provide the ultimate level of fast, accurate and safe analysis.

You can determine flash points up to 400 °C using either standard mode for where the flash point is known or search mode for when it is not. Once complete, the PM-93 graphically displays results on the LCD screen. You can save results to the internal memory, email the results, export results to a USB, or transfer results via QR code.

SQC software analyses results, calculates mean, standard deviation, standard error and repeatability and plots results in accordance with ASTM D6299.

Methods

ASTM D93, ASTM D8175, IP 34, BS EN ISO 2719 & BS 2000-34, SW-846 1010B

Features

  • Easy operation with touch screen
  • Automatic flash point determination
  • Gas or electric ignition
  • Patented ‘SafeFlash’ auto fire extinguisher
  • Large capacity memory for storage of 30 test profiles, operator names and test methods
  • Real time display of test progress and storage for 2000 results
  • LIMS connectivity, results can be emailed, saved to USB or transferred via QR code
  • Statistical Quality Control (SQC) software to analyse results to conform to ASTM D6299

Details

Direct correlation to ASTM D613 (ISO 5165) engine tests:
Single action lifting arm operation
Intuitive user menu with colour touch screen
Fully automatic test procedure, little operator intervention required

Enhanced test throughput:
Unique test profiles to help improve efficiency when testing a large number of samples
Fast warm up and high capacity air cooling minimises cycle time

Precision and accuracy:
Automatic flash point ensures test repeatability
Results are corrected for variations in atmospheric pressure using the integral barometer
Unique alignment design ensures lid and cup locate perfectly
Password protected ‘calibration’ mode allows verification of instrument
SQC enhanced software to ensure precision and accuracy

Supplied with: SETA-CRM Pensky-Martens Flash Point (3 pack),cup (fitted), allen key 0.89 mm, allen key 1.50 mm, mains cable, quick start guide and USB with the instrument handbook.

NOTE 1): The ‘SafeFlash’ fire extinguisher is a factory fit accessory and complies with ISO 2719. It must be ordered at the time of purchase or the unit must be returned to a service centre for installation. Comprises extinguisher, manifold assembly, regulator, fire sensor and remote operating button.

Specifications

Ignition systemElectric or gas
Flash detectionThermal
CoolingForced air (integral fan)
Heating RateFast heating rate (>10 °C/min) and standard; 5.5 °C/min; 3 °C/min; 1.3 °C/min; 1 °C/min
Temperature rangeAmbient +5 °C to 400 °C
Internal Memory30 programmable test profiles, operator names, test methods and unlimited tests results
Fire DetectionThermal with optional integral Fire Extinguisher
Size/Weight38.5 x 24 x 50 cm / 25 kg
Voltage230 Vac +/-15% or 110 Vac +/-15%. Frequency 50/60 Hz. Auto-sensing – no selector switch required

Spares and Accessories