Seta Oil Test Centrifuge – 4 place – 90000-3
The Seta Oil Test Centrifuge is a heated chamber centrifuge used to determine water and sediment in oils. The large diameter chamber, with four place rotor and cushioned adaptors allows the use of 6 and 8 inch conical, pear-shaped, trace sediment and 12.5 ml tubes.
The swing out rotor ensures that samples are held near vertical when at rest allowing precise measurement of sediment. The centrifuge features a hinged lid that opens to the fully upright position to allow easy loading and unloading of tubes between tests.
Speed, temperature and timing are microprocessor controlled. Run parameters are entered via a liquid proof ‘touch panel’ and digital display.
View accessories required for each test method
Methods
ASTM D91, ASTM D893, ASTM D1290, ASTM D1796, ASTM D2273, ASTM D2709, ASTM D2711, ASTM D4007, ISO 3734, ISO 9030, ISO 15301
Features
- Isothermal heaters fitted internally to the centrifuge chamber
- Temperature control using unique algorithm to compensate for air resistance heating
- 4 place swing out rotor
- Static, near vertical bucket positioning
- Specifically designed for testing oils
- Heated chamber, ambient to 80 °C
- Pre-heat facility to ensure bowl is at test temperature prior to test
- Universal buckets for 12.5 ml, 6 and 8 inch glassware
- Microprocessor control with 99 memories
- Optional trolley
Details
Precise heating for accurate results
Isothermal heaters fitted internally to the centrifuge chamber give precise temperature control to within +/-1 °C.
Fast and easy operation
9 programme memories allow a user to configure the instrument to automatically operate to the specified test parameters, saving time, easing operation and assisting repeatability. A ‘Quick Run’ mode allows short runs of up to 99 seconds.
Safety features
A motor driven lid latch with a safety interlock prevents the centrifuge operation while the lid is unlocked. Imbalance detection and over-temperature protections protects the motor and chamber from damage.
Setafuge Hazard Reduction System
The optional Setafuge Hazard Reduction System (90010-0) can be fitted into the centrifuge to extract vapours from the chamber and significantly reduce vapour hazards.
Supplied with;
Rotor
Buckets and cushions
Specifications
Maximum Relative Centrifugal Force (RCF) | 2210 g |
Maximum Rotational Speed | 3000 rpm |
Temperature Range | 20 to 80 °C ±1 °C |
Timer Range | 10 seconds to 1 hour (10 second steps) 1 hour to 99 hours 59 minutes (1 minute steps) Continuous running until manually stopped |
Control | Microprocessor with 99 memory capacity |
Capacity | 4 Samples |
Voltage range | 230 V, 50/60 Hz (Optional transformer for 110/120 V) |
Power | 950 W |
Size (HxWxD) | 410 x 560 x 640 mm |
Weight | 72.5 kg |