Setaflash Series 3e Flash Point Tester - 30020-0

Setaflash Series 3e Flash Point Tester – 30020-0

The Setaflash Series 3e closed cup flash point tester is a compact, bench top/portable instrument used to carry out ‘flash/no flash’ tests and determine flash point temperatures.

The Series 3e can perform tests at temperatures ranging from 0 to 300 °C (with the 13870-0 Seta Cooling Module allowing temperatures below ambient).

Operated by a touchscreen, icons guide you through the simple test procedure, with a test taking less than 2 minutes with just 2 or 4 ml of sample required. Up to 10,000 test results can be saved in the instruments internal memory which can be saved and transferred via USB port for data analysis.

Methods

ASTM D3278, ASTM D3828, ASTM D7236, ASTM D8174, IP 523IP 524, IP 534, CLP Regulations EC No 1272/2008, EPA 1020 B, ISO 3679, ISO 3680 (Obs), SW-846 1020C

Features

  • Electric ignition
  • Ambient to 300 °C
  • Small sample size
  • Under 2 minute test
  • Simple operation
  • Compact, rugged, portable design
  • Suitable for unknown samples using ramp mode
  • Automated flash detection
  • Automatic barometric pressure correction
  • Colour touchscreen
  • USB port and result storage

Details

Ease of use:
All functions on the Series 3 including test parameters, instrument status and test results are accessed via the touchscreen display and push button. An audible prompt sounds when a flash is achieved and an automatic flash detector enhances repeatability.

Calibration and verification:
Calibration of the Series 3 can be performed in the field, eliminating the costs and time associated with sending the instrument to a service centre.

Service and maintenance:
The rugged design ensures minimal operational maintenance in needed.

Supplied with: 2 ml syringe, ignitor, Allen key, mains lead, quick start guide and instruction manual

Specifications

Temperature rangeAmbient to 300 °C
Temperature ramp modeAutomatic or custom settings (ramp rate: up to 6 °C/min and beep rate: up to 5 °C/B intervals) Extended test range, 100 °C from start temperature
Temperature selection °C or °F
User interfaceColour LCD display – touch screen sensitive
NavigationControl knob (rotational and push-down)
Data storage1GB (internal memory and integrated USB port)
Sample size2 ml or 4 ml
Pressure correctionAutomatic pressure correction
Voltage110 to 250 V a.c~50/60 Hz (universal)
Power200 W (Max.)
Size (HxWxD)19.5 cm x 29.5 cm x 14 cm