VariDens – Density Meter
The VariDens benchtop density meter offers precise and reliable measurements. With the proven oscillating U-tube principle, the VariDens provides density and concentration measurements of pharmaceutical products, chemicals, and petrochemicals.
The VariDens density meter offers an optional humidity and air pressure sensor, and precise temperature control ensures extended use even under difficult ambient conditions. Available accessories and interchangeable fittings reduce the installation effort and save time when evaluating products from other manufacturers.
The modular design allows the VariDens to be combined with other VariFamily instruments setting a new standard for density measurement instruments.
Features
- Precise and fast measurement
- Integrated reference oscillator
- Fast temperature control
- Automatic detection of gas bubbles
- Modular connectivity to your internet browser and tablet or device of your choice
- An intuitive user interface, multiple users
- Easy cleaning
- Compliant with 21 CFR part 11
- ELN/LIMS ready
Details
The SCHMIDT+HAENSCH VariDens Density Meter is a rugged, sealed system designed for continuous density measurement in a broad range of applications with the flexibility to handle any measuring inquiry in even the harshest environments.
Intuitive, user-friendly software and the large, color-TFT touchscreen interface reduces the setup times for protocols.
The VariDens Density Meter can be combined with other SCHMIDT+HAENSCH laboratory instruments enabling lab managers to handle and sign off data while lab technicians are still measuring.
Specifications
VariDens C | VariDens B | |
Measuring range | 0 – 3 g/cm3 | 0 – 3 g/cm3 |
Temperature range | 10 – 95°C* (regulated Peltier thermoelement) | 10 – 95°C* (regulated Peltier thermoelement) |
Pressure range | 0 to 10 Bar (0 to 145 psi) | 0 to 10 Bar (0 to 145 psi) |
Density accuracy | ± 0.0001 g/cm3 | ± 0.00005 g/cm3 |
Reproducibility | ± 0.0001 g/cm3 | ± 0.00002 g/cm3 |
Resolution | 0.0001 g/cm3 | 0.00001 g/cm3 |
Temperature accuracy | ± 0.05 °C | ± 0.03 °C |
Reproducibility | ± 0.02 °C | ± 0.01 °C |
Resolution | 0.01 °C | 0.01 °C |
Temperature stabilization time | 25 °C to 20 °C, ± 0.01 °C in 18 minutes* | 25 °C to 20 °C, ± 0.01 °C in 18 minutes* |
Wetted parts | Borosilicate glass, PTFE | Borosilicate glass, PTFE |
Bubble detection | Yes (combined: visual-electronic) | Yes (combined: visual-electronic) |
Drying | Integrated air pump | Integrated air pump |
Temperature compensation | Yes | Yes |
Minimum sample volume | 1.5 ml | 1.5 ml |
Measuring time per sample** | < 10 sec | < 30 sec |
Display | 7″ touchscreen***, 800 x 480 Pixel | 7″ touchscreen***, 800 x 480 Pixel |
Interfaces | Ethernet 10/100/1000, RS232, USB, WLAN*** | Ethernet 10/100/1000, RS232, USB, WLAN*** |
Ambient conditions | Temperature 10 °C to 35 °C, Humidity < 85 % not condensing | Temperature 10 °C to 35 °C, Humidity < 85 % not condensing |
Power | VAC 90-265; 50/60 Hz; 150 W max | VAC 90-265; 50/60 Hz; 150 W max |
Conformity | International Pharmacopoeia, ASTM, AOAC, DIN, 21 CFR part 11, ICUMSA and others | International Pharmacopoeia, ASTM, AOAC, DIN, 21 CFR part 11, ICUMSA and others |
Dimensions | With display: 430 x 300 x 235 mm (w x d x h) / without display: 430 x 300 x 150 mm (w x d x h) | With display: 430 x 300 x 235 mm (w x d x h) / without display: 430 x 300 x 150 mm (w x d x h) |
Highlights | Modularity; user friendliness; fast temperature control; easy cleaning; virtually maintenance free product, 21 CFR part 11 conform; ELN; LIMS ready | Modularity; user friendliness; fast temperature control; easy cleaning; virtually maintenance free product, 21 CFR part 11 conform; ELN; LIMS ready |
*Valid for ambient temp. = 20 °C
** After temperature stabilisation
*** Optional